Table 2 The transmitted signal and the intraclass correlation between IAPdot vs. IAPivp

From: Feasibility and accuracy of continuous intraabdominal pressure monitoring with a capsular device in human pilot trial

 

Transmitted data pair

Continuously IAP monitoring achievement

Single measure ICC

95% Confidence interval

Case 1

182

yes

0.9255

0.8451 ~ 0.9578

Case 2

170

yes

0.9123

0.8831 ~ 0.9344

Case 3

143

yes

0.9124

0.8740 ~ 0.9383

Case 4

244

yes

0.9678

0.9215 ~ 0.9830

Case 5

462

yes

0.9360

0.8658 ~ 0.9632

Case 6

320

yes

0.8791

0.5300 ~ 0.9482

summarized

1521

yes

0.9119

0.8821 ~ 0.9320